Details on MP-RAGE

 

 

MP-RAGE (Magnetization Prepared RApid Gradient Echo) is a rapid gradient-echo technique in which a preparation pulse (or pulses) is applied before the acquisition sequence to enhance contrast.  [John P. Mugler III and James R. Brookeman, “Three-Dimensional Magnetization-Prepared Rapid Gradient-Echo Imaging (3D MP RAGE)”, Magnetic Resonance in Medicine 15, pp.  152--157 (1990)]

 

In our sequence a single inversion pulse is followed by a 400 ms delay time. As a result our MP-RAGE images are T1-weighted.  The excitation pulse has a 10 degree flip angle, TE is 4 ms, and TR is 10 ms.  The resolution is 128 x 256 x 256. The corresponding dimensions are approximately 160 mm by 250 mm by 250 mm.  (The precise voxel dimensions for each volume are, as always, included in the corresponding header.ascii file). There is one acquisition. Total acquisition time is 8 minutes and 55 seconds.

 

 

 Return to RIRE home page